Introduction
The TechMize TH2848 series is a high-precision LCR meter manufactured by Changzhou Tonghui Electronic Co. Ltd. These meters are designed for measuring impedance parameters with exceptional accuracy, stability, and speed. Featuring auto-balance bridge technology and multiple testing configurations, the TH2848 series is ideal for evaluating passive components, semiconductors, dielectric materials, and more.
Key Features
-
Wide Frequency Range: 4Hz - 10MHz
-
High Accuracy: Basic accuracy of 0.05%
-
Auto Balance Bridge Technology
-
Multiple Measurement Modes: Spot measurement, list sweep, and graphic sweep
-
Advanced Display: 10.1-inch capacitive touch screen with 1280x800 resolution
-
High Power Output: 20VAC/100mAAC with built-in DC bias (±40VDC / 100mADC)
-
Fast Testing: Dual CPU architecture enabling test speeds up to 400 times per second (2.5ms)
-
User-Friendly Interface: Linux-based operating system with touch operation and embedded help
-
Comprehensive Measurement Capabilities: Four-parameter measurement, graphical analysis, and powerful sorting functions
-
Compatibility: Supports SCPI/MODBUS command set, compatible with KEYSIGHT E4980A, E4980AL, and HP4284A
-
Integrated Testing Functions: Piezoelectric conductivity circle test, dielectric constant evaluation, and more
-
Enhanced Connectivity: USB, LAN, GPIB, RS232C, RS485, and Handler interface support
Applications
The TechMize TH2848 series is widely used in various industries for analyzing and evaluating components, materials, and devices:
-
Passive Components: Inductors, capacitors, resistors, transformers, crystals, network components
-
Semiconductor Components: LED driver ICs, varactor diodes, transistors, integrated circuits
-
Other Components: PCBs, relays, switches, cables, batteries
-
Dielectric & Magnetic Materials: Plastics, ceramics, ferrites, amorphous materials
-
Semiconductor & Liquid Crystal Materials: Electrical conductivity, dielectric constant analysis
-
Piezoelectric Devices: Resonance frequency, anti-resonance frequency, mechanical coupling coefficient testing
Analogues
Several European manufacturers produce LCR meters with similar specifications to the TechMize TH2840 series:
-
Hameg (Rohde & Schwarz) HM8118: Frequency range up to 200 kHz, basic accuracy of 0.05%.
-
Wayne Kerr 6500B: Frequency range up to 5 MHz, basic accuracy of 0.05%.
-
Hioki IM3536: Frequency range up to 3 GHz, basic accuracy of 0.05%.
These models offer high measurement accuracy and a broad frequency range, making them viable alternatives to the TechMize TH2848 series.

Technical Specifications
| Product Model | TH2848-02L | TH2848-02 | TH2848-05 | TH2848-10 | |
| Display | Monitor | 10.1-inch (diagonal) capacitive touch screen | |||
| Scale | 16:9 | ||||
| Resolution | 1280×RGB×800 | ||||
| Measured Parameters | Methods | Four parameters can be selected arbitrarily | |||
| AC | Cp,Cs,Lp,Ls,Rp,Rs,|Z|,|Y|,R,X,G,B,θ,D,Q,VAC,IAC, | ||||
| DC | RDC | ||||
| Piezoelectricity | Ct,Dt,Fs,Fp,Fp-Fs,Zmin,Zmax,F1,F2,F2-F1,Gmax,C0,C1,R1,L,Kp,Keff,Kt,K31,K33,Qm,ε,εr | ||||
| Dielectric | Cp,D,ε,|ε|,εr',εr'',tanδ,Q | ||||
| Test Frequency | Range | 4Hz-2MHz | 4Hz-2MHz | 4Hz-5MHz | 4Hz-10MHz |
| Accuracy | 0.01% | ||||
| Resolution | 0.1mHz 4.0000Hz-99.9999Hz | ||||
| 1mHz 100.000Hz-999.999Hz | |||||
| 10mHz 1.00000kHz-9.99999kHz | |||||
| 100mHz 10.0000kHz-99.9999kHz | |||||
| 1Hz 100.000kHz-999.999kHz | |||||
| 10Hz 1.00000MHz-9.99999MHz | |||||
| AC Test Signal Mode | Rating Value (ALC OFF) | The set voltage is the Hcur voltage when the test terminal is open-circuit. The set current is the current from Hcur when the test terminal is short-circuited. |
|||
| Constant Value(ALC ON) | Keep the voltage on the DUT the same as the set value Keep the current on the DUT the same as the set value |
||||
| Test level | Voltage range | 0Vrms - 2Vrms | 0mVrms-20Vrms F<=1MHz | ||
| 0mVrms-15Vrms 1MHz<F<=2MHz | |||||
| 0mVrms-2Vrms 2MHz<F<=MHz | |||||
| 0mVrms-1Vrms 5MHz<F<=10MHz | |||||
| Accuracy | ±(10%×set value+2mV)(AC<=Vrms) | ||||
| ±(10%×set value+5mV)(AC>2Vrms) | |||||
| Resolution | 0.1mVrms 0mVrms-0.2Vrms | ||||
| 0.2mVrms 0.2Vrms-0.5Vrms | |||||
| 0.5mVrms 0.5Vrms-1Vrms | |||||
| 1mVrms 1Vrms-10Vrms | |||||
| 10mVrms 10Vrms-20Vrms | |||||
| Current range | 0mArms-20mArms | 0mArms-100mArms | |||
| Resolution (100Ω internal resistance) | 1μArms 0Arms-2mArms | ||||
| 2μArms 2mArms-5mArms | |||||
| 5μArms 5mArms-10mArms | |||||
| 10μArms 10mArms-100mArms | |||||
| RDC test | Voltage range | 100mV-1V | |||
| Resolution | 100μV | ||||
| Current range | 0mA-10mA | ||||
| Resolution | 10μA | ||||
| DC Bias | voltage range | 0V-±10V | 0V-±40V | ||
| Accuracy | 1%×set voltage+5mV AC<=2V | ||||
| 2%×set voltage+8mV AC>2V | |||||
| Resolution | 0.1mV 0V - ±5V | ||||
| 1mV ±5V - ±40V | |||||
| Current range | 0mA - ±100mA | ||||
| Resolution | 1μA 0mA-50mA | ||||
| 10μA 50mA-100mA | |||||
| Voltage source | voltage range | -10V - 10V | |||
| Resolution | 1mV | ||||
| Current range | -45mA - +45mA | ||||
| Output Impedance | 100Ω | ||||
| Cylinder control | Switching control | ||||
| Turn on | Turns on within the set time from 0-60s | ||||
| Turn off | Turns off within the set time from 0-60s | ||||
| Test Side Configuration | Four-terminal pair | ||||
| Test cable length | 0m,1m | ||||
| Output Impedance | 100Ω,±1%@1kHz | ||||
| Mathematical operation | Absolute deviation Δ from nominal value, percentage deviation Δ% from nominal value | ||||
| Equivalence mode | Series, parallel | ||||
| Calibration function | OPEN,SHORT,LOAD | ||||
| Measured average | 1-255 times | ||||
| Range Selection | AUTO,HOLD | ||||
| Trigger Mode | Continuous, single | ||||
| Trigger delay | 0-60s | ||||
| specific function | One-click screenshot, one-click record, embedded help system | ||||
| Range Configurat | LCR | 100mΩ,1Ω,10Ω,20Ω,50Ω,100Ω,200Ω,500Ω,1kΩ,2kΩ,5kΩ,10kΩ,20kΩ,50kΩ,100kΩ | |||
| RDC | 10Ω,20Ω,50Ω,100Ω,200Ω,500Ω,1kΩ,2kΩ,5kΩ,10kΩ,20kΩ,50kΩ,100kΩ | ||||
| Measuring time (ms/time) (Frequency >= 100kHz) | Fast: 2.5ms Medium: 90ms Slow: 220ms | ||||
| Maximum accuracy | 0.05%(refer to specifications) | ||||
| Measurement display range | |||||
| Cs,Cp | 0.00001pF - 9.99999F | ||||
| Ls,Lp | 0.00001μH - 99.9999kH | ||||
| D | 0.00001 - 9.99999 | ||||
| Q | 0.00001 - 99999.9 | ||||
| R,Rs,Rp,X,Z,RDC | 0.001mΩ - 99.9999MΩ | ||||
| G,B,Y | 0.00001μs - 99.9999S | ||||
| VDC | ±0V - ±999.9999V | ||||
| IDC | ±0A - ±999.9999A | ||||
| θr | -6.28318 | ||||
| θd | -179.999° - 179.999° | ||||
| Δ% | ±(0.000% - 999.9%) | ||||
| Multifunction parameter list scanning | Points | 201 points, average can be set for each point, each point can be sorted individually | |||
| Parameters | Test frequency, AC voltage, AC current, DC BIAS voltage, DC BIAS current, with dielectric constant test function based on the parameter of each point in this list | ||||
| Trigger Mode | Sequential SEQ: When triggered once, measurements are taken at all scan points, /EOM/INDEX is output only once Step STEP: performs one scan point measurement per trigger, outputs /EOM/INDEX for each point, but list scan comparator result is only output at the last /EOM |
||||
| Other features | 1.Multiple copy functions for both scanning and test parameters 2.Time delay can be set for each scanning point |
||||
| Dielectric constant | Dielectric Material Testing Solutions | ||||
| Comparator | Each scanning point can measure up to four test parameters, each parameter can be set upper and lower limits, all test parameters are qualified to output PASS signal, otherwise output FAIL signal, no judgment if no upper and lower limits are set. | ||||
| Graphic Scanning | Scanning Points | Points 51, 101, 201, 401, 801 are optional | |||
| Results Display | Extreme values for each parameter and the scanned parameter value at the point where the cursor is located with the corresponding test parameter value | ||||
| Piezoelectric testing | Piezoelectric device and piezoelectric material testing solutions, integrated piezoelectric admittance circle function | ||||
| Scanning track | 1-4 test parameters can be selected arbitrarily, and the scanning curve can be divided into one screen, two screens and four screens. | ||||
| Display range | Real-time automatic, locked | ||||
| Coordinate scale | Logarithmic, linear | ||||
| Scanning parameters | Frequency, AC Voltage, AC Current, DCV BIAS/DCI BIAS | ||||
| Trigger method, Single | Trigger manually once, one scan from start to finish is completed, the next trigger signal starts a new scan | ||||
| Trigger method, Sequential | Infinite loop scanning from start to finish | ||||
| Results Saving | Graphics, documentation | ||||
| Comparator | Bin Staging | 10Bin,PASS,FAIL | |||
| Bin deviation setting | Deviation value, percent deviation value, off | ||||
| Bin mode | Tolerance, continuous | ||||
| Bin Count | 0-99999 | ||||
| BIN Judgement | Up to four parameters can be set for the limit range. If the results of the four test parameters fall within the set range, the corresponding BIN number is displayed. If it exceeds the maximum BIN number range set, it displays FAIL. Test parameters without upper and lower limits set will be automatically ignored for BIN judgement. | ||||
| PASS/FAIL Indication | If it meets Bin1-10 criteria, the PASS light on the front panel lights up; otherwise, the FAIL light is illuminated. | ||||
| Data Cache | 201 measurement results can be batch-read. | ||||
| Storage | Internal | The instrument has 8GB of built-in storage space, after removing the system occupancy, the user can use about 6GB of space. | |||
| External USB | Test setup files, screenshot graphics, record files | ||||
| Keyboard Lock | Lockable front panel keys | ||||
| Interfaces | USB HOST | 2 USB HOST ports, can be connected to the mouse, keyboard, U disk can only be used one at a time | |||
| USB DEVICE | Universal Serial Bus socket, small Class B (4 contact positions); compliant with USB TMC-USB 488 and USB2.0, female connector is used to connect an external controller. | ||||
| LAN | 10/100M Ethernet, 8-pin, two speeds adaptive | ||||
| HANDLER | For Bin Staging Signal Output | ||||
| GPIB | Standard | ||||
| RS232C | Standard 9-pin, Crossed | ||||
| RS485 | Standard | ||||
| Power-on warm-up time | 60 minutes. | ||||
| Input Voltage | 100-120VAC/198-242VAC selectable,47-63Hz | ||||
| Power supply power | Not less than 130VA | ||||
| Dimensions(WxHxD)mm | 430x177x265 | ||||
| Weights | 11kg | ||||
Accessary
| Standard | ||||||
| Accessories name | Model | |||||
| Four-end clamp with card | TH26005D | ![]() |
||||
| Short circuit | TH26010 | ![]() |
||||
| Kelvin test lead with box four-terminal insulation and lock | TH26011BS | ![]() |
||||
| Optional | ||||||
| Accessories name | Model | |||||
| Box patch test cable | TH26009B | ![]() |
||||
| Magnetic ring fixture | TH26008A | ![]() |
||||
| Dielectric test fixture | TH26077 | ![]() |
||||
Conclusion
The TechMize TH2848 series of LCR meters is a powerful, precise, and versatile solution for component and material testing across various industries. With high-speed performance, extensive measurement capabilities, and advanced features, it is an excellent choice for professionals seeking accurate impedance analysis and evaluation.